    I-I-I    Selamat datang di website Sistem Penelitian dan Pengabmas Terpadu (SIPANDU) Pusat Penelitian dan Pengabdian Kepada Masyarakat (PPPM) Poltekkes Kemenkes Semarang    I-I-I    Anda dapat mendownload full text jurnal penelitian dan pengabdian masyarakat jika anda terdaftar sebagai dosen dan mahasiswa Poltekkes Kemenkes Semarang, semoga bermanfaat untuk ilmu yang bermanfaat    I-I-I    

01 September 2012 oleh SITI DARYATI. S.Si., M.Sc , ; ;
LINK ISSN: 1829-5754, Vol. 8, No. 3, September 2012

Pengaruh Variasi Milliampere-Seconds (mAs) Terhadap Entrance Skin Exposure Dan Gambaran Anatomi Radiografi Pada Pemeriksaan Columna Vertebra Lumbal Proyeksi Ap

Abstract

Quantitative Descriptive research with experimental opproach was done with x-ray machiine type UD150l-30E. The data of experiment was taken with made eight radiographs and the ESE measurement is conducted using TLD which layed on surface of abdomen phantom or at central point. Exposure was taken with eight times with eight variation of mAs. It is about 8,10,12,14,16,18,20 and 22 mAs with 80 kV. The result shown that there is a effetct of variation mAs to ESE and quality of radiographic anatomy image. Higher mAs result higher ESE and increase mAs about two interval reslut average value is 142 mR. Variation of mAs result variation of image quality. A good image quality with 19,1 cm part thickness resulted with use 18 mAs and 20 mAs with 80 kV. Optimum mAs with 80 kV in this research’s is 18 mAs.

Key Word: variation mAs, Entrance Skin Exposure (ESE), quality of radiographic anatomy image, columna vertebra lumbal